Pre-Calibrated Room Setup
3. TARGET-FREE SCANNING SETUP
Concrete Structure
Requirement:
1, Keep away from the vibration source, otherwise vibration-proof
foundation should be required to consider.
2, Room Temperature Requirement: 20±5 ℃.
3, If room temperature is unstable, please make sure that room
temperature difference for encode acquisition and workpiece scanning
shall not exceed ±5℃.
4, Concrete surface roughness ≤ Ra3.2 (otherwise encode targets &
positioning targets will fall easily).