ROMER Absolute Arm presentation
ROMER ABSOLUTE ARM Manual V2.3.1 (29/05/2012) 7
A
Environment
The product is suitable for use in an atmosphere appropriate for permanent human habitation.
The product is not suitable for use in harsh or explosive environments.
WARNING
The person in charge of the product should contact local safety authorities and safety experts
before working in any hazardous areas or in close proximity to electrical installations.
AC power supplies are suitable for use in dry environments only and not under adverse
conditions.
-
Mechanical safety
CAUTION
Mechanical closing locations
Danger of pinch points
While the risk of personal injury is low, for your personal safety, avoid handling the areas
of the device shown in the following diagram while operating the device
NOTICE
Arm installation
Fall or drop of the device
The base of the ROMER ABSOLUTE ARM is designed to provide a reliable measurement
platform when placed on a level and stable work surface. The ROMER Mounting Ring also
can be fixed on a plate to a rigid support via the fixing holes.
Always ensure that the arm is correctly and strongly fixed to the base
Always ensure that the base is correctly and strongly fixed to the support
NOTICE
Arm use
Accuracy deterioration
Before any use, always ensure nothing adverse happened on the arm (accuracy
verification, counterbalance pressure, buttons response)
When moving the arm, always pay attention to obstacles that may damage the device.
The arm and the computer should be put in their respective cases when not in use or
during transportation to avoid any damage that may affect the calibration
Moreover, for any transport by plane, boat, train or truck, the product should be packed
in foam for protection, with shock detectors
Do not modify counterbalance pressure or strength as this will affect the arm calibration