ROMER ABSOLUTE ARM Manual V2.3.1 (29/05/2012) 97
J
Accuracy specifications J.1.2
Version 1
Version 2
Counterbalance with hard stop.
Counterbalance with locking system.
ROMER Absolute Arm accuracies
The following data represent the accuracies of the different sizes and levels of ROMER
Absolute Arm.
All the values refer to the ASME B89.4.22 standard and are displayed in millimetres.
All probing specifications are achieved with a ROMER Absolute Arm mounted on a base plate
or magnetic base plate and using a 15mm steel ball probe with a length of 50mm under stable
environment conditions.
Volumetric Accuracy
Single Point Repeatability
73
75
73
75
V1
V2
V1
V2
V1
V2
V1
V2
1500
0.037
0.037
0.025
0.025
2000 (6ft)
0.042
0.042
0.023
0.023
0.030
0.030
0.016
0.016
2500 (8ft)
0.051
0.051
0.029
0.029
0.038
0.038
0.020
0.020
3000 (10ft)
0.095
0.075
0.049
0.044
0.065
0.059
0.033
0.030
3500 (12ft)
0.130
0.100
0.061
0.057
0.095
0.079
0.043
0.040
4000
0.149
0.125
0.075
0.069
0.120
0.099
0.061
0.055
4500
0.170
0.150
0.082
0.082
0.150
0.120
0.070
0.070
Tube inspection arm accuracy specifications J.1.3
ROMER Absolute Arm accuracies
The following data represent the accuracies of the different sizes and levels of ROMER
Absolute Arm when used in Arm Tube configuration (stronger gas pump).
All the values refer to the ASME B89.4.22 standard and are displayed in millimetres.
All probing specifications are achieved with a ROMER Absolute Arm mounted on a base plate
or magnetic base plate and using a 15mm steel ball probe with a length of 50mm under stable
environment conditions.
Volumetric Accuracy
Single Point Repeatability
73
73
2500 (8ft)
0.059
0.044
3000 (10ft)
0.086
0.068

Table of Contents