DT4261 DIGITAL MULTIMETER Remote Operation Manual
P8/11
Command
Explanation
:SYST:BEEP
Set the action of the beeper.
Syntax :SYST:BEEP <data×1>
<data> “0” (off) or “1” (on)
Example; “:SYST:BEEP 1”
Response “OK” or “CMD ERR”
:SYST:BLIT
Set the backlight.
Syntax :SYST:BLIT <data×1>
<data> “0” (off) or “1” (on)
Example; “:SYST:BLIT 1”
Response “OK” or “CMD ERR”
:SYST:BLA
Set the auto backlight.
Syntax :SYST: BLA <data×1>
<data> “0” (off) or “1” (on)
Example; “:SYST: BLA 1”
Response “OK” or “CMD ERR”
:SYST:BATT?
Query the battery level.
Syntax :SYST:BATT?
Response <data×1>
<data> Four steps remaining battery level.
“3” or “2” or “1” or “0”
:SYST:ZEROADJ
Execute zero adjustment.
Syntax :SYST:ZEROADJ
Response “OK”
or “CMD ERR” (Fail, or unsupported function)
:SYST:FILTER
Set the filter.
Syntax :SYST: FILTER <data×2>
<data1> “0” (off) or “1” (on)
<data2> 100” (100Hz) or 500 (500Hz)
Example; “:SYST: FILTER 1,100
Response “OK” or “CMD ERR”
:MEAS:AUTOV?
Query the status DC or AC of AutoV function.
Syntax :MEAS: AUTOV?
Response 0 (DC) or “1 (AC)
or EXE ERR (Other than AutoV or LoZV function)
:SYST:INIT
Put the meter to power-on-reset state.
Syntax :SYST: INIT
Response “OK”