Replacing Batteries or Fuse
2.1 Replacing Batteries or Fuse
WARNING
To avoid electric shock, turn off the MEASURE key,
disconnect the test leads from the measuring object
before replacing the battery cover.
After replacing the batteries, reattach the cover and
secure the screw before using the instrument.
Battery may explode if mistreated. Do not
shortcircuit, recharge, disassemble or dispose of in
re.
Replace the fuse only with one of the specied type,
characteristics, rated current, and rated voltage. Do
not use fuses other than those specied (especially,
do not use a fuse with higher-rated current) or do
not short circuit and use the fuse holder. Doing so
may damage the instrument and result in bodily
injury.
Fuse type: FF0.5 AH/1000 V (70 172 40.0.500:
SIBA GmbH) (Fast blow, arc-extinguishing material
included, and high breaking capacity)
The fuses can be purchased via authorized Hioki
distributor or reseller. (Fuse replacement not
required for the IR4053 or IR4055)
To prevent instrument damage or electric shock,
use only the screw for securing the battery cover
in place that are originally installed. If you have lost
a screw or nd that a screw is damaged, please
contact your authorized Hioki distributor or reseller.
HIOKI IR4057B961-00