IR4053
IR4055
IR4056
IR4057-50
IR4053-10
IR4055-11
IR4056-20
IR4056-21
IR4053
IR4055
IR4056
IR4057-50
IR4053-10
IR4055-11
IR4056-20
IR4056-21
EN
EN

Instruction Manual
Dec. 2020 Edition 1
IR4057B961-00 20-12H
(IR4052C960-00)
INSULATION TESTER

Instruction Manual
Dec. 2020 Edition 1
IR4057B961-00 20-12H
(IR4052C960-00)
INSULATION TESTER
HIOKI IR4057B961-00

Questions and Answers

  • M
    Maria MorganSep 3, 2025
    How to stabilize measured values on Hioki IR4053 Test Equipment?
    • M
      Manuel GordonSep 3, 2025
      If the measured values from your Hioki Test Equipment vary and fail to stabilize, disconnect the circuit breaker for any nearby charging circuits. If this is not possible, use the lowest measured value as the measurement result.
  • L
    Lee WhiteSep 18, 2025
    What does Err5 01 mean on my Hioki IR4053 Test Equipment?
    • M
      Michael RussellSep 18, 2025
      If your Hioki Test Equipment displays 'Err5 01' indicating an abnormality in the measurement circuit, replace the batteries. If there is no apparent improvement, the instrument needs repair.
  • L
    Laura LambertSep 16, 2025
    What does Err 0ADJ mean on my Hioki IR4053 Test Equipment?
    • T
      timdavisSep 16, 2025
      If your Hioki Test Equipment displays 'Err 0?ADJ' indicating that zero adjustment is outside the allowable range during low resistance measurement, verify that there is no broken connection in the test leads. Zero adjustment can be performed for readings of up to 3 ?. Ensure that the wiring resistance is 3 ? or less.
  • A
    Andrea RossSep 14, 2025
    What does Err4 mean on my Hioki IR4053?
    • L
      Lauren MooreSep 14, 2025
      If your Hioki Test Equipment displays 'Err4' indicating an abnormality in the measurement circuit, replace the batteries. If there is no apparent improvement, the instrument needs repair.
  • K
    Kyle JacksonSep 12, 2025
    What type of fuse should I use in my Hioki IR4053 Test Equipment?
    • R
      richardsonjonathanSep 12, 2025
      If the specified fuse has not been installed in your Hioki Test Equipment, use only the specified type of fuse.
  • D
    Daniel BeckSep 12, 2025
    How to resolve Err1 on Hioki IR4053 Test Equipment during zero adjustment?
    • A
      Ashley SmithSep 12, 2025
      If your Hioki Test Equipment shows 'Err1' and is unable to perform zero adjustment during low resistance measurement, verify that there is no broken connection in the test leads. Zero adjustment can be performed for readings of up to 3 ?. Ensure that the wiring resistance is 3 ? or less.
  • C
    Chad RodriguezSep 9, 2025
    What to do if the insulation resistance range falls outside device specifications during Hioki IR4053 Test Equipment calibration?
    • J
      joseph51Sep 9, 2025
      When calibrating the Hioki Test Equipment, if the accuracy of the insulation resistance range falls outside the device specifications, use the test leads that came with the instrument or its optional test leads to perform the calibration procedure. If the issue persists, replace the deteriorated test lead with a new one.
  • R
    Ryan HillSep 7, 2025
    Why does Hioki IR4053 Test Equipment give different measured values each time?
    • M
      Mrs. Rachel ArnoldSep 7, 2025
      If you get a different measured value each time you measure the same object with your Hioki Test Equipment, allow an adequate amount of time (about 1 hour to 1 day) to pass after the first measurement before repeating measurement. Also, measure the object under the same temperature and humidity conditions.
  • A
    Angela ReedSep 5, 2025
    Why do measured values fluctuate over time on Hioki IR4053 during insulation resistance measurement?
    • M
      Melissa PachecoSep 5, 2025
      When measured values acquired through insulation resistance measurement or PV measurement fluctuate over time on your Hioki Test Equipment, if it is possible to remove the capacitor, do so. If not, use the lowest measured value as the measurement result. Alternatively, select measured values acquired one minute after the start of the measurement, or after they have stabilized if a large capacitance component is included.
  • J
    james20Aug 20, 2025
    What to do if I am unable to perform measurement on my Hioki Test Equipment?
    • S
      sarahbrownAug 21, 2025
      If you are unable to perform a measurement with your Hioki Test Equipment, first turn off the MEASURE key and then press it again. Also, ensure that the measuring object is separated from all power sources before you attempt to measure it.

Hioki IR4053 Specifications

General IconGeneral
Measurement Voltage50V, 125V, 250V, 500V, 1000V
CategoryInsulation Tester
Insulation Resistance Range (1000V)0.001 MΩ to 4000 MΩ
AC Voltage Measurement0 to 600 V AC
DisplayLCD
Power SupplyAA alkaline battery (LR6) × 4
Dimensions159 mm × 177 mm × 53 mm
Weight680 g