IR3455
HIGH VOLTAGE
INSULATION TESTER
Instruction Manual
Oct. 2016 Edition 1
IR3455A961-00 16-10H
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

Table of Contents

Questions and Answers

  • T
    Thomas BurgessAug 2, 2025
    What to do if Hioki IR3455 Test Equipment power fails when measuring insulation resistance?
    • M
      Mr. Michael HarrisAug 2, 2025
      If your Hioki Test Equipment loses power while measuring insulation resistance, there are several potential causes: * The battery power might be low, in which case you should replace it with a new one. * The battery pack may need charging; ensure it is fully charged. * The GUARD terminal might be short-circuited with the test lead connected to the + terminal. Verify the connection to the test lead clips.
  • S
    stevensheltonAug 7, 2025
    How to fix Hioki IR3455 not turning on?
    • M
      Megan PriceAug 7, 2025
      If your Hioki Test Equipment is not turning on, consider the following: * Ensure a battery is installed. If not, install a new battery. * If the battery power is low, replace it with a new battery. * Verify that the battery polarity is correct. * If using a battery pack, ensure it is charged. Charge the battery pack if necessary. * Confirm that the battery selector switch is in the correct position.
  • S
    Sean DavisAug 12, 2025
    Why is my Hioki IR3455 battery pack not charging?
    • E
      Edward JohnsonAug 12, 2025
      If the battery pack of your Hioki Test Equipment is not charging, check these two things: * Ensure the power plug of the AC adapter is fully inserted. * Confirm that the battery pack is properly installed. If not, install the battery pack.
  • J
    Joseph LarsonAug 17, 2025
    Why is my Hioki IR3455 Test Equipment resistance measurement value incorrect?
    • M
      Megan WilliamsAug 17, 2025
      An incorrect resistance measurement value on your Hioki Test Equipment can stem from a few issues: * The test lead may be damaged, in which case you should replace it. * The test lead might not be fully inserted; ensure it is. * The test leads might be connected to the wrong terminals; double-check the terminals.
  • A
    Andrew CastroAug 21, 2025
    Why is monitored voltage low during resistance measurement with Hioki Test Equipment?
    • B
      Brenda GloverAug 21, 2025
      The output voltage is intentionally lowered for measurement of low resistance values.
  • N
    nicholas01Aug 27, 2025
    Why resistance is not measured in temperature compensation mode on Hioki IR3455 Test Equipment?
    • L
      Lisa ThomasAug 27, 2025
      To measure resistance in temperature compensation mode using your Hioki Test Equipment, you must measure temperature before resistance.

Hioki IR3455 Specifications

General IconGeneral
CategoryInsulation Resistance Tester
Operating temperature and humidity-10°C to 50°C, 85% RH or less (no condensation)
Measurement Range (Insulation)0.01 MΩ to 1000 GΩ
Resolution0.01 MΩ
DisplayDigital Display
Power SupplyAA batteries (LR6)
Weightincluding batteries
Test Voltage250 V, 500 V, 1000 V