7.1 General Specifications
1
2
3
4
5
6
7
8
9
10
11
付録
索引
141
Specifications
7
7.1 General Specifications
Operating
temperature
and humidity
-10°C to 40°C (14°F to 104°F), less than
80% RH (no condensation)
40°C to 50°C (104°F to 122°F), at 50°C
and below relative with linear decrease
up to 50% RH
(Battery pack charge: 0°C to 40°C, less
than 80% RH)
Storage
temperature
and humidity
-10°C to 50°C (14°F to 122°F), less than
90% RH (no condensation)
Battery pack: -20°C to 30°C, less than
80% RH (no condensation)
Guaranteed
accuracy
period
1 year
Operating
environment
Indoors, Pollution degree 2,
Up to 2000 m (6562 ft.) ASL
Measuring
method
DC voltage application (insulation resis-
tance) and mean-value rectification (voltage)
A-D
conversion
Double integral
Display
LCD with backlight displaying up to a
count of 999
Overflow
indication
>, OF
Underflow
indication
<, -OF
Display update
rate
Insulation resistance/leakage current:
Once/sec. (0.25 times/sec. if averaging
function used)
Output voltage monitor: Twice/sec.
Voltage measurement: Fourth/sec.
Temperature measurement:Once/sec.
Bar graph: Twice/sec.