4.4 Step Voltage Test
98
The voltage is increased when one minute has passed at
each voltage step. When 5 minutes has passed in total,
measurement automatically stops.
The voltage application time is adjustable.
See 6.2 "Changing and Checking Voltage Application Time for Step
Voltage Test" (page 128).
The voltage application time cannot be varied for each
step.
4.4.1 Setting and Conducting a Step Voltage
Test
Procedure
1. Press the
key in standby state,
and the voltage indication will blink.
2. Press the key to choose
[STEP2.50 kVSET] or [STEP5.00
kVSET].
The voltage value will advance rapidly if
the key is held down.
Choosing [
5.00 kVSET
] with the key
and then pressing the key is a short-
cut to select STEP.
STEP5.00 kV
1 kV
2 kV
3 kV
4 kV
5 kV
1 min. 2 min. 3 min. 4 min. 5 min.
Measurement
(1)
Measurement
(2)
Measurement
(3)
Measurement
(4)
Measurement
(5)
End of
Measurement