8.1 Troubleshooting
1
2
3
4
5
6
7
8
9
10
11
付録
索引
155
If the cause is unknown, try resetting the system.
See 8.4 "Performing System Reset" (page 158).
Monitored
voltage during
resistance
measurement
is low.
Is the resis-
tance small?
The output
voltage is low-
ered for mea-
surement of
low resistance
values.
Appendix 1
(P.163)
Temperature
is not
measured.
Is the sensor
inserted fully?
Insert the sen-
sor fully.
2.5 (P.58)
Resistance is
not measured
in temperature
compensation
mode.
Have you mea-
sured tempera-
ture first?
Measure tem-
perature
before resis-
tance.
4.3 (P.93)
The
instrument
cannot
communicate
with the PC.
Is the USB
cable connec-
tor inserted
fully?
Insert the USB
cable connec-
tor fully.
6.4 (P.136)
Power fails
upon
measuring
insulation
resistance.
Is the battery
power low?
Replace with
new battery.
2.1.1
(P.36)
Is the battery
pack charged?
Charge the
battery pack.
2.1.4
(P.47)
Is the GUARD
terminal short-
circuited with
the test lead
connected to
the + terminal?
Check the
connection to
the test lead
clips.
3.2.1
Proce-
dure 3.
(P.65)
Problem Check Item Action
Reference
Section