EasyManua.ls Logo

Hioki IR3455 - Temperature Measurement

Hioki IR3455
175 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.2 Measurement Specifications
1
2
3
4
5
6
7
8
9
10
11
付録
索引
151
7.2.4 Temperature Measurement
Temperature and humidity range for guaranteed accuracy:
23±5°C less than 80% RH (no condensation)
Accuracy when using with the 9631 TEMPERATURE SEN-
SOR
When the 9631-05 TEMPERATURE SENSOR is used, the accuracy is
guaranteed within 0.0°C to 40.0°C.
Measuring range Measurement accuracy
-10.0°C to -0.1°C
±1.5°C
0.0°C to 40.0°C
±1.0°C
40.1°C to 70.0°C
±1.5°C
Measurement Range, Accuracy
Temperature
characteristics
Measurement accuracy × 0.5 is added to
the measurement accuracy. (when the
ambient temperature is not 23±5°C)
Response time
Approx. 100 sec. Including the period of
time for the response of the 9631-01,
9631-05 TEMPERATURE SENSORs.
(Reference value: Period of time until
90% of the change in temperature is
reflected in the indication)
Influence of
radioactive RF
electromagnet-
ic field
±2°C at 3V/m

Table of Contents