2.1.2 Installing the Battery Pack
(Rechargeable nickel-hydrogen bat-
tery)
Use the optional 9459 BATTERY PACK. The operating
time is longer than that with alkaline batteries, and the
pack is rechargeable.
Battery pack is dispatched in an uncharged state. Charge
before use.
Procedure
See 2.1.4 "Charging the Battery Pack" (page 47).
For battery operation, use only the Hioki
Model 9459 BATTERY PACK. We do not
take any responsibility for accidents or
damage related to the use of any other
batteries.
To avoid heat buildup, rupture, or leak-
age of the battery, do not use if dam-
aged, wires are exposed, or the battery/
instrument connector is damaged.
To avoid electric shock, be sure to dis-
connect the test leads from the instru-
ment, turn off power, and disconnect
the AC adapter from the instrument,
before installing or removing the battery
pack.
Battery may explode if mistreated. Do
not short-circuit, disassemble or dis-
pose of in fire. Do not recharge alkaline
batteries. Handle and dispose of batter-
ies in accordance with local regulations.
Take care not to step on the battery pack
power cable, as this may damage it.