8.4 Performing System Reset
158
System reset returns the settings of the instrument to their
defaults (excluding date and time), but this will not clear the
memory data.
Procedure
1. While holding down the key in
standby state, press the key.
[rESEt] appears.
2. Press the key, and [rESEt] will
blink and the LCD returns to the standby
screen. System reset is complete.
The table below shows the default settings.
8.4 Performing System Reset
Setting Items Settings
Resistance/current
Resistance
Test voltage
250 V
Timer
OFF
PI interval
t1=1 min., t2=10 min.
Temperature
compensation
OFF
Table No. displayed first
when temperature
compensation is selected.
0
Reference temperature for
temperature compensation
20°C for table No. 0 to 8
40°C for table No. 9
Step voltage test
OFF
Duration of one step in
step voltage test
1 min.
Logging recording interval
1 min.
Average
OFF
Auto power off
ON