6.2 Changing and Checking Voltage Application Time for
Step Voltage Test
128
• Change the voltage application time for step voltage test.
Selectable presets: 30 sec., 1 min., 2 min. , 5 min.
(Default is 1 min.)
• The voltage application time to set up is the application
time for a voltage step, not the total application time for 5
steps.
6.2.1 Changing Time Setting
Procedure
1. Press the key while in standby
state, and the voltage indication will
blink.
2. Press the key to choose
[STEP2.50 kVSET] or [STEP5.00
kVSET].
• If the key is held down the voltage value
changes rapidly.
• Choosing [
5.00 kVSET
] with the
key and then pressing the key is a
shortcut to select STEP.
6.2 Changing and Checking
Voltage Application Time for
Step Voltage Test