Questions and Answers

  • T
    tonyadelgadoSep 2, 2025
    Why does the MEASURE key enable the lock function after releasing it in Hioki IR4053?
    • R
      rdiazSep 2, 2025
      If, in the 500 V, 1000 V, or PV range, turning on the MEASURE key enables the lock function that has been released on your Hioki Test Equipment, replace the batteries with new alkaline batteries and release the lock function again.
  • Z
    zstevensonAug 31, 2025
    What to do if the MEASURE key is not working for the lead with a switch on my Hioki IR4053 Test Equipment?
    • A
      Anthony OrtizAug 31, 2025
      If the MEASURE key is not working for the lead with a switch on your Hioki Test Equipment, ensure that you fully insert the plug to the limit without any gaps.
  • T
    Thomas BarberAug 29, 2025
    How to fix Hioki IR4053 when unable to perform measurement in 500 V, 1000 V or PV range?
    • J
      Joseph HartAug 29, 2025
      If you are unable to perform measurements in the 500 V, 1000 V, or PV range on your Hioki Test Equipment, release the lock function.
  • A
    Ana CarterAug 27, 2025
    What should I do if the protective fuse has tripped in my Hioki IR4053 Test Equipment?
    • P
      Pamela WilliamsonAug 28, 2025
      If the protective fuse in your Hioki Test Equipment has tripped, replace it with the indicated fuse.
  • R
    Ruth BaileyAug 26, 2025
    What to do if my Hioki IR4053 shows the maximum display value?
    • R
      Rita AdamsAug 26, 2025
      If the measured value on your Hioki Test Equipment is shown as the maximum display value, check the continuity of the test lead with a tester. Also, check the connection between the test leads and the instrument, and the connection of the tips of the test leads.
  • A
    Andrew DouglasAug 25, 2025
    Why do the batteries in my Hioki IR4053 run out so quickly?
    • E
      Elizabeth RiveraAug 25, 2025
      If the batteries in your Hioki Test Equipment run out immediately, replace them with alkaline batteries.
  • D
    David GarciaAug 22, 2025
    How to fix Hioki IR4053 when the power is not turning on?
    • N
      nataliezimmermanAug 22, 2025
      If your Hioki Test Equipment is not turning on, replace the batteries with new ones and ensure they are installed in the correct orientation.

Hioki IR4053 Specifications

General IconGeneral
Measurement Voltage50V, 125V, 250V, 500V, 1000V
CategoryInsulation Tester
Insulation Resistance Range (1000V)0.001 MΩ to 4000 MΩ
AC Voltage Measurement0 to 600 V AC
DisplayLCD
Power SupplyAA alkaline battery (LR6) × 4
Dimensions159 mm × 177 mm × 53 mm
Weight680 g