A-15
State Comment: Returned as the device state if the C-arm microprocessor returned an
unrecognized response to the fast query.
Action: This may be due to a communications error between the AT and the C-arm.
Follow the procedure under "BAD CHECKSUM IN MESSAGE." to diagnose I/O
errors. Otherwise, report this error to Hologic Technical Support.
Comment: For the state machine driver, this error indicates that either the protocol
attempted to execute a command that is not supported; that an instruction that requires
a register reference did not provide one; or, more commonly, an attempt was made to
access the external data buffers and they were either not available or were still being
using by the application.
Action: If you are running any program other than standard Hologic software, remove
it and retry the scan (e.g., do not run a scan while connected to a network or attempt to
run a scan under Windows). Otherwise, report this error to Hologic Technical Support.
Motor_xx Comment: For a stepper motor, the application attempted to move a motor in the
opposite direction in which it is currently moving (i.e., a motor's motion can be
extended in the same direction as the current motion but it cannot reverse motion).
Action: Report this error to Hologic Technical Support.
Invalid parameter(s) [E_INVALID_PARAM_I] Comment: The value of the parameter associated with a device command was out of
range.
Action: Report the application in error to Hologic Technical Support.
Invalid serial number in ARRC.TXT. Action: This is not an error if the machine is not yet calibrated.
Invalid serial number in environ.bat file. Action: Machine serial number is not valid. Call Hologic Technical Support.
Invalid values in bad detector map Action: See instructions for running hardware checking program.