A-22
User
or
Software
When restoring files from optical ("Optical" on main menu), you selected an archive
that contains no scans.
Action: Report this problem to Hologic Technical Support.
There are no records in the Normals Curve database.
Software The reference curve database appears to be totally empty. This condition should never
occur in normal operations.
Action: The Hologic software may not have been installed correctly.
The X-ray controller is not generating A/C line interrupts [E_NO_TIMER_INTERRUPT_I]
Hardware The X-ray Controller assembly generates an interrupt every 1/120 of a second (60HZ
line) or every 1/100 of a second (50Hz line). The PC clock generates 18 interrupts a
second (regardless of line frequency). If two clock ticks occur with NO Motor & Drive
board interrupts between them, then this message is produced.
Action: Report this error to Hologic Technical Support. Include as much information as
possible about what you were doing at the time.
Refer to the “NO A/C Line Interrupts” message at start-up on page 6-9.
Truncated Patient Record ... Hardware
or
Software
This message indicate problems performing I/O operations (usually to the hard drive,
but possibly to diskette or optical).
Action: Ensure that there is adequate space on the hard drive (Check the "Room for ..."
message in the status window. You should always leave room for 1 or 2 scans). There
may be a disk problem; appropriate diagnostics should be run. If you are unable to find
a hardware problem, the message should be reported to Hologic Technical Support.
Include as much information as possible about what you were doing at the time.
device: Unable To Allocate Resources
Software This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
ERROR MESSAGE
HARDWARE
/SOFTWARE
ERROR
LOCUS COMMENT/ACTION