A-14
Invalid address for message packet [E_INVALID_ADDR_I] Comment: The address of a message packet was invalid. The MAPI driver keeps an
address table that maps each byte address to its respective communications link (SSI,
SCI or C/C). The selected address was not assigned to any link and therefore was not
recognized by the MAPI driver.
Action: Report this error to Hologic Technical Support.
Invalid flattening data [E_INVALID_FLAT_DATA_S] Comment: The most likely cause is that the signal for one or more of the detectors is
higher than allowed.
Action: One or more detectors is bad. See instructions for running hardware checking
program.
Invalid identification file. Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Invalid machine type in ARRC.TXT. Action: This is not an error if the machine is not yet calibrated.
Invalid mode for device driver [E_INVALID_MODE_I] C-arm Comment: Returned by the C-arm if an application attempts to change the X-ray mode
when X-rays are already on.
Action: Report this error to Hologic Technical Support.
ERROR MESSAGE
HARDWARE
/SOFTWARE
ERROR
LOCUS COMMENT/ACTION