A-26
Unexpected message from remote device [E_UNEXPECTED_RESPONSE_I] Comment: The first character in a fast query response was not the fast query command
("?").
Action: Follow the procedure under "BAD CHECKSUM IN MESSAGE." for isolating
communications bus I/O errors.
Unknown ROI type in DrawROI Software These messages should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
device: Unrecognized Command Code
Software This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Unrecognized device command [E_UNKNOWN_COMMAND_I] Comment: Returned by all devices in response to an unrecognized command.
Action: Report this error to Hologic Technical Support. Include as much information as
possible about what you were doing at the time.
Unsolicited message from C/C [E_UNSOLICITED_MESSAGE_I] Action: Report this error to Hologic Technical Support. Include as much information as
possible about what you were doing at the time.
X-ray controller is not generating A/C line interrupts [E_NO_TIMER_INTERRUPT_I]
DAS
Global
A request was made to read the a/d's and there was no system line frequency interrupts
(required to synchronize DAS with C-arm).
The system was on-line and another on-line command was issued but there were still
no timer interrupts.
Action: Check that the Instrument Power switch and the X-Ray Enable key are on.
Also, see the discussion of E_INTERLOCK_INHIBIT_I, E_EMERGENCY_STOP_I
and E_COLLISION_IMMINENT_I.
Refer to the “NO A/C Line Interrupts” message at start-up on page 6-9.
ERROR MESSAGE
HARDWARE
/SOFTWARE
ERROR
LOCUS COMMENT/ACTION