Specifications and Model Number
7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual
2-9
Note 1: Customer must supply tagging information; Up to 3 lines allowed.
(22 characters for each line)
Note 2: This option for use with HPW7000 assembly only.
Note 3: Standard range should be selected if measuring the following
conductivity, resistivity or TDS ranges:
STANDARD RANGE
Conductivity TDS Cell Constant
0-1.999 0-1999 ppb 0.01
0-19.99 0-19,990 ppb 0.1
0-199.9 0-199.9 ppm 1.0
0-1999 0-1999 ppm 10
0-19,990 0-10,000 ppm 25 or 50
Resistivity Cell Constant
0-19.99M 0.01
0-1.999M 0.1
0-199.9k 1.0
0-19.99k 10
0-1.999k 25 or 50
Note 4: Wide range should be selected if measuring the following
conductivity, TDS or weight % concentration ranges:
WIDE RANGE
Conductivity TDS Cell Constant
100-1999 0-1999 ppm 0.1
1000-19,990 0-19,990 ppm 1.0
10-199.9 mS 0-199.9 ppt 10
10-500 mS 25
20-1000 mS 50
Weight % Concentration Cell Constant
0-18.00% Hydrochloric Acid 50
0-19.99% Sulfuric Acid 50
0-19.99% Sodium Hydroxide 25 or 50
0-19.99% Sodium Chloride 25 or 50