Honeywell 9782 Series - Table 5-1 Procedure for Performing I;O Setup

Honeywell 9782 Series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Conductivity/Resistivity Analyzer/Controller
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/995-6
Table 5-1 Procedure for Performing I/O Setup
Step Screen Action
1
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV | NEXT
Go to I/O SETUP.
2
RELAY SETTINGS
RLY1 RLY2 RLY3 RLY4
ACLN RNGS ALM3 ALM4
DIAGNOSTICS
| PREV | NEXT
The first screen shows the present use of available relays.
This information is read-only.
In this example the 9782 has four relays. Relay 1 is used
for AutoClean. Relay 2 is used for indication of automatic
output range switching. Relay 3 is used to trigger an alarm
annunciator for Alarm 3. Relay 4 is used for Alarm 4
annunciation.
Go to the next screen.
3
ANALOG OUTPUTS
OUTPUT1 C1
OUTPUT2 C2
OUTPUT3 RATO
DIAGNOSTICS
| PREV | NEXT
The next screen shows the present assignments for the
optional current or voltage outputs, if available. This
information is read-only.
1
Go to the next screen.
4
ATTENTION
RECONFIG OF ANY I/O
SETUP PARM WILL
RESET CONFIG PARMS
TO FACTORY DEFAULTS
MENU TO EXIT
DIAGNOSTICS
| PREV | NEXT
I/O setup is normally performed only once, before field
connections are made to the Analyzer/Controller. You can
also use this menu to view parameters. If you do not
change anything, you will not cause a cold reset of the
unit. You can exit now using the MENU key. If you go
past this “Attention” screen, you must go through the entire
sequence of I/O setup screens before returning to the
main menu.
Go to the next screen.
5
ADVANCED FEATURES
AUTOCLEAN YES
RANGE STATUS NO
DIAGNOSTICS
| | NEXT
Specify whether AutoClean should be enabled. If YES, a
relay will be dedicated to this use.
Specify whether you want to use a relay to indicate when
the range shifts for the analog output based on Cell 1’s
input. (Output scaling is specified in Step 13.) If YES, a
relay will be assigned to indicate when the Cell 1 range
switch occurs.
Go to the next screen.
1
If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show “DE
ENABLED”, instead of displaying output assignments.