28
APT4000PH
Configuration
In the Configuration mode you set the analyzer parameters.
Activate with CONF
Enter mode code “1200”
Edit parameter with and ,
confirm/continue with ENTER.
(End with CONF ENTER.)
During configuration the
APT4000PH remains in the Hold
mode for reasons of safety. The
output current is frozen (at its last
value or at a preset fixed value,
depending on the configuration),
limit and alarm contacts are inac-
tive.
The controller is in the configured
state, Sensoface is off, mode indi-
cator “Configuration” is on.
Activate
Hold
Input errors
The configuration parameters are
checked during the input. In the
case of an incorrect input “Err” is
displayed for approx. 3 s. The
incorrect parameters cannot be
stored. Input must be repeated.
End
End with CONF. The measured
value and Hold are displayed alter-
nately, “enter” flashes. End Hold
mode with ENTER.
The display shows the measured
value. The output current remains
frozen for another 20 s (Hold icon
on, “hourglass” flashes).
Hold icon
29
Menu group
Code
Output 1
o1.
Output 2
o2.
Temperature
compensation
tc.
Calibration mode
CA.
Alarm
settings
AL.
Relay / Controller
rL.
Rinsing and cali-
bration probes
Pb.
Display
Select menu group
Previous
menu group:
Menu structure of configuration
Menu item 1
Menu item 2
Menu item ...
The configuration steps are assigned to different menu groups.
With the arrow keys you can jump between the individual menu
groups.
Each menu group contains menu items for setting the parameters.
Pressing ENTER opens a menu item. The values are edited using the
arrow keys. Pressing ENTER confirms/stores the settings.
Return to measurement: Press CONF.
Select menu item