Calibration error messages
Asymmetry potential out of range (±60 mV)
Electrode worn out
Buffer solution contaminated
Buffer does not belong to configured buffer set
Temperature probe not immersed in buffer
solution (for automatic temperature compensation)
Wrong buffer temperature selected
(for manual temperature specification)
Nominal electrode zero point pH 7
Symbol flashes:
Problem
Possible causes
Problems during recognition
of the buffer solution
Same or similar buffer solution was used
for both calibration steps
Buffer solution used does not belong to
buffer set currently configured in the analyzer
Electrode slope out of range
(80-103 %)
Electrode worn out
Buffer solution contaminated
Buffer does not belong to configured
buffer set
Temperature probe not immersed in
buffer solution (for automatic
temperature compensation)
Wrong buffer temperature set (for manual
temperature specification)
Electrode used has different nominal slope
89
Problems during recognition
of the buffer solution (continued)
During manual calibration the buffer solutions
were not used in the specified order
Buffer solutions contaminated
Wrong buffer temperature set
(for manual temperature specification)
Electrode defective
Electrode not connected
Electrode cable defective
Symbol flashes:
Problem
Possible causes
Calibration was canceled after approx.
2 min, because the electrode drift was too
large.
Electrode defective
Electrode dirty
No electrolyte in the electrode
Electrode cable insufficiently shielded or defective
Strong electric fields influence the measurement
Major temperature fluctuation of the buffer solution
No buffer solution or extremely diluted