EasyManua.ls Logo

Honeywell APT4000PH - Page 23

Honeywell APT4000PH
59 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
APT4000PH
Configuration
Output 2
Temperature error. Output current during Hold.
Select °C/°F
Select temperature probe
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA for temp error
Hold mode
End:
Press CONF, then ENTER
Output 2
o2.
Code
Display
Menu group
Select menu item
45
OFF
(ON)
Code
Display
Action
Choices
o2.
22 mA signal for error message
Select with key
Proceed with ENTER
Output signal during Hold
LAST:
During Hold the last measured value
is maintained at the output
FIX:
During Hold a value (to be entered)
is maintained at the output
Select with key
Proceed with ENTER
LAST
(FIX)
Only with FIX selected:
Enter current which is to flow at the
output during Hold
Select with key,
edit number with key,
proceed with ENTER
021.0 mA
(000.0 to
021.0 mA)
Output signal during Hold:
Output current
[mA]
Output signal Hold
Setting FIX = 21.0 mA
Hold active
21
4
Output signal Hold
Setting LAST
Hold active

Related product manuals