45
OFF
(ON)
Code
Display
Action
Choices
o2.
22 mA signal for error message
Select with key
Proceed with ENTER
Output signal during Hold
LAST:
During Hold the last measured value
is maintained at the output
FIX:
During Hold a value (to be entered)
is maintained at the output
Select with key
Proceed with ENTER
LAST
(FIX)
Only with FIX selected:
Enter current which is to flow at the
output during Hold
Select with key,
edit number with key,
proceed with ENTER
021.0 mA
(000.0 to
021.0 mA)
Output signal during Hold:
Output current
[mA]
Output signal Hold
Setting FIX = 21.0 mA
Hold active
21
4
Output signal Hold
Setting LAST
Hold active