Midas
®
Gas Detector
7-4
7.1.5 Review Calibration ‘ CAL
1. Select the review calibration ‘ CAL’ submenu
and press ‘3’ to accept.
2. The number of days remaining to the next
calibration due date is displayed. If the number
of days is set to zero then no further reminders
will be displayed.
3. Press 3 to display the number of days until
cartridge expiration.
4. Press 3 to display the year and press 3
again to display the month and day of the last
calibration.
5. Press ‘3’ to return to step 1.
6. Repeat steps 1 through 4 to view the settings
again or press ‘s’ up or ‘t’ down to select
another submenu.
7. Press ‘X’ to return to normal operation.
7.1.6 Review Date and Time ‘timE’
Select the review date and time ‘timE’ submenu and
press ‘3’ to accept.
1. The current year setting is displayed.
2. Press ‘3’ to display the current month and day
3. Press ‘3’ again to display the current time.
4. Press ‘3’ to return to step 1.
5. Repeat steps 1 through 5 to view the settings
again or press ‘s’ up or ‘t’ down to select
another submenu.
6. Press ‘X’ to return to normal operation.

Table of Contents

Other manuals for Honeywell LonWorks

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell LonWorks and is the answer not in the manual?

Honeywell LonWorks Specifications

General IconGeneral
BrandHoneywell
ModelLonWorks
CategoryGas Detectors
LanguageEnglish

Related product manuals