Release 1 34-ST-25-19 RMA 3000 Remote Meter Assemblies User Manual 41
6/04
IEC Temperature
Classification
Equipment intended for installation directly within the hazardous location must also
be classified for the maximum surface temperature that can be generated under
normal or fault conditions as referenced to the maximum operating ambient of the
equipment. The maximum surface temperature must be less than the minimum auto-
ignition temperature of the hazardous atmosphere present. The temperature shall be
indicated in identification numbers as listed in the following table.
Maximum Temperature
Degrees C Degrees F Identification Number
450 842 T1
300 572 T2
200 392 T3
135 275 T4
100 212 T5
85 185 T6
Certification and
Conformity Details
Table 8 CENELEC / LCIE Certification
Code Description
3D
Flameproof, Supply 45 Vdc, IP 66/67EEx d IIC T6.
3A
Intrinsically Safe EEx ia IIC T5, 40 Ta 93°C.
Flameproof, Supply 45 Vdc, IP 66/67 EEx d IIC T6.
LCIE Intrinsic Safety Parameters
(1)
U
i
= 30 V
I
i
= 100 mA
P
i
= 1.2 W
C
i
= 4.2 nF
R
i
= 0
L
i
= 0 With no integral indicator, or with
integral Smart Meter, option SM.
L
i
=
150 µH
With Analog Meter, option ME.
(1)
Install in accordance with Honeywell drawing 51204243.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell RMA 3000 and is the answer not in the manual?

Honeywell RMA 3000 Specifications

General IconGeneral
BrandHoneywell
ModelRMA 3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals