58 Chapter 1 D82008753M 10/13
New VSI Controller
1 flash: The battery needs charging or there is a bad
connection to the battery. Try charging the battery.
2 flashes: The left hand motor has a bad connection.
3 flashes: The left hand motor has a short circuit to a
battery connection.
4 flashes: The right hand motor has a bad connection.
5 flashes: The right hand motor has a short circuit to a
battery connection.
7 flashes: A joystick fault is indicted. Make sure that the
joystick is in the center position before switching
on the controller.
8 flashes: A controller fault is indicated.
9 flashes: Problem with the brake circuit. Check that the
manual release levers are in the drive position.
10 flashes: An excessive voltage has been applied to the
controller. This is usually caused by a poor
battery connection.
Speed Bar Cycling: Controller locked.
Speed Gauge Cycling: Check if battery charger is still
connected to controller or controller fault.