Loading...
IEC 1010-1:1990+A1 / EN 61010-1:1993
UL 3111
CSA-C22.2 No.1010.1:1993
This Product meets the requirement of the European Communities (EC)
EMC Directive 89/336/EEC.
Less than 60 dBA
EN55011/CISPR 11 (ISM, Group 1, Class A equipment),
IEC 555-2 and IEC 555-3
EN50082-1 Code Notes
IEC 801-2 (ESD) 8kV AD
IEC 801-3 (Rad.) 3 V/m
IEC 801-4 (EFT) 1kV
2
1
1
Safety
Product Regulations
EMC
Sound Pressure
Level
Emissions
Immunity
Performance Codes:
1 PASS - Normal operation, no effect.
2 PASS - Temporary degradation, self recoverable.
3 PASS - Temporary degradation, operator intervention required.
4 FAIL - Not recoverable, component damage.
Enviromental
Conditions
Definitions
Temperature
Humidity
Power
Installation category (overvoltage category) Signal level, special equipment or parts of
equipment, telecommunication, electronic etc., with smaller transient overvoltages than
installation (overvoltage category)
Installation category (overvoltage category) Local level, appliances, portable
equipment etc., with smaller transient overvoltages than installation category
Indoor use only.
Altitude up to 3000 m. (10,000 ft.)
I:
II.
II:
III.
Instrument - 0 degrees C to 50 degrees C (32 degrees F to 122 degrees F)
Disk Media - 10 degrees C to 40 degrees C (50 degrees F to 104 degrees F)
Probes/cables - 0 degrees C to 65 degrees C (32 degrees F to 149 degrees F)
Relative humidity 8 to 80% at 40 degrees C (104 degrees F)
CAT , Pollution degree 2
HP 16700A : ~Line 115/230 volts ± 20%, 48-66 Hz, 610 Watts max.
II
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com