17preg
Product Regulations
Safety
EMC
Sound Pressure
Level
IEC 348:1978 / HD 401 S1:1981
UL 1244
CSA-C22.2 No. 231 (Series M-89)
This Product meets the requirements of the European Communities (EC)
EMC Directive 89/336/EEC.
(none)
N/A
Emissions
Immunity
EN55011/CISPR 11 (ISM, Group 1, Class A equipment)
EN50082-1 Code Notes
IEC 555-2 1
IEC 555-3 1
IEC 801-2 (ESD) 8kV AD 2
IEC 801-3 (Rad.) 3V/m 1
IEC 801-4 (EFT) 1kV 1
Performance Codes:
1 Pass - Normal operation, no effect.
2 Pass - Temporary degradation, self recoverable.
3 Pass - Temporary degradation, operator intervention required.
4 Fail - Not recoverable, component damage.
Notes:
Enviromental
Conditions
Definitions
Temperature
Humidity
Power
Installation category (overvoltage category) Signal level, special equipment or parts of
equipment, telecommunication, electronic etc., with smaller transient overvoltages than
installation (overvoltage category)
Installation category (overvoltage category) Local level, appliances, portable
equipment etc., with smaller transient overvoltages than installation category
Indoor use only.
Altitude up to 3000 m. (10,000 ft.)
I:
II.
II:
III.
Instrument - 0 degrees C to 50 degrees C (32 degrees F to 122 degrees F)
Disk Media - 10 degrees C to 40 degrees C (50 degrees F to 104 degrees F)
Probes/cables - 0 degrees C to 65 degrees C (32 degrees F to 149 degrees F)
Relative humidity 8 to 80% at 40 degrees C (104 degrees F)
(From host frame.)