The
thermal
stability
of
the
measurement
environment
is
also
a
very
important
consideration.
If
you
are
making
measurements
in
a
stable
temperature
environ-
ment,
disabling
the
autozero
function
can
nearly
double
the
reading
rate
with
almost
no
loss
in
accuracy.
Any
range
or
function
change
that
takes
place
is
automatically
accompanied
by
an
autozero
update
which
removes
any
ac-
cumulated
offsets.
Other
Factors
Ir
sncing
the
Reading
Rate
1.
You
can
speed
the
reading
rate
by
selecting
a
fixed
range
instead
of
allowing
the
multimeter
to
autorange
every
reading.
2.
AC
voltage
measurements
have
a
built-in
600mS
settling
time.
Resistance
measurements
can
be
made
as
fast
as
DC
voltage
measurements
except
on
the
3Mohm
and
30Mohm
ranges
where
a
settling
time
(20mS
and
200mS
respective-
ly)
is
needed
for
accurate
measurements.
3.
The
maximum
possible
reading
rate
is
with
3
1/2
digits
displayed,
autozero
off,
any
DC
voltage
function,
60
Hz
power
selected,
manual
range,
a
positive
voltage
measured,
and
internal
trigger
selected.
4.
It
is
important
to
note
that
even
though
the
3468A
may
be
taking
readings
at
a
faster
rate,
the
display
is
only
updated
at
a
maximum
rate
of
five
times
per
second.
Faster
reading
rates
are
intended
for
and
only
accessible
through
HP-IL.