8662A Option HI
1
08662-92019
Page 4 of 20
On page 3-80, Table 3-12, Initializing; Response to Special Function 00, add the
following to the end of the Table:
Parameter
10 MHz Reference Source
Condition
Internal 10 MHz Reference
On page 3-82, Table 3-14, Special Function Status Codes, change entry 51 to read as
follows:
51
External 10 MHz Reference Enabled
SECTION IV. PERFORMANCE TESTS
On page
4-23,
in Section
4-11,
INTERNAL TIME BASE AGING TEST, change step 1
of the PROCEDURE to read as follows:
1.
Press the Blue Key, SPECIAL, and 00 to initialize the instrument and to select
the internal 10 MHz reference oscillator.
SECTION V. ADJUSTMENTS
On page 5-9, in Paragraph 5-7, 10 MHz REFERENCE OSCILLATOR ADJUSTMENT,
change step 4 under PROCEDURE to read as follows:
4.
Press the Blue Key, Special, and 00 to select the internal 10 MHz reference
oscillator.