EasyManua.ls Logo

HP 8753D Option 011 - Page 34

HP 8753D Option 011
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HP 8753D
Performance
Test
Record (14
of 18)
For
300
kHz|3
GHz
Analyzers
Hewlett-Pac
kard
Company
Mo
del HP
8753D Option
011
Rep
ort Num
ber
Serial
Num
b
er
Date
7
7
12.
Receiv
er
Magnitude
Dynamic
Accuracy
G F
j
G
0
F
j
T
est
P
ort
Input
P
o
wer
8496A
A
ttn.
(dB)
T
est
P
ort
Measuremen
t
(dBm)
Exp
ected
Measuremen
t
(dBm)
Dynamic
Accuracy
(Calculated)
Sp
ec.
(dB)
Meas.
Uncer.
(dB)
Channel
B
0
10
0
0.075
6
0.008
0
20
(Ref
)
10 Reference Reference Reference Reference Reference
0
30
20
0.050
6
0.008
0
40
30
0.050
6
0.008
0
50
40
0.050
6
0.008
0
60
50
0.058
6
0.017
0
70
60
0.089
6
0.017
0
80
70
0.240
6
0.017
0
90
80
0.680
6
0.017
0
100
90
1.950
6
0.027
Channel
A
0
10
0
0.075
6
0.008
0
20
(Ref
)
10 Reference Reference Reference Reference Reference
0
30
20
0.050
6
0.008
0
40
30
0.050
6
0.008
0
50
40
0.050
6
0.008
0
60 50
0.058
6
0.017
0
70
60
0.089
6
0.017
0
80
70
0.240
6
0.017
0
90
80
0.680
6
0.017
0
100
90
1.950
6
0.027
DRAFT
3/21/106 15:21
Performance Test Record 2a-15
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Related product manuals