HT65
EN - 20
AC TRMS Current with transducer clamp
Range Output ratio Resolution
Accuracy (*)
(50Hz
1kHz)
Overload protection
1000mA 1V/1A 1mA
r(2.5%rdg + 10dgt)
1000VDC/ACrms
10A
100mV/1A
0.01A
30A
40A (**)
10mV/1A
r(3.5%rdg + 30dgt) (***)
100A
0.1A
r(2.5%rdg + 10dgt)
300A
400A (**)
1mV/1A
r(3.5%rdg + 30dgt) (***)
1000A
1A
r(2.5%rdg + 10dgt)
3000A
(*) Accuracy referred to only instrument without transducer clamp; Accuracy specified from 5% to 100% of the measuring range
(**) With HT4006 transducer; (***) Accuracy instrument + clamp
For not sinusoidal waveforms the accuracy is: r(10.0%rdg + 10dgt)
Resistance and Continuity test
Range (*) Resolution Accuracy Buzzer Overload protection
400.0: 0.1: r(1.2%rdg + 4dgt)
<50:
250VDC/ACrms
4.000k: 0.001k: r(1.0%rdg + 2dgt)
40.00k: 0.01k:
r(1.2%rdg + 2dgt)
400.0k: 0.1k:
4.000M: 0.001M:
40.00M: 0.01M: r(2.0%rdg + 3dgt)
Frequency (electrical circuits)
Range (*) Resolution Accuracy Overload protection
10Hz y 10kHz
0.001Hz÷0.01kHz
r(1.5%rdg + 5dgt)
600VDC/ACrms
Frequency (electronic circuits)
Range Resolution Accuracy Overload protection
9.999Hz 0.001Hz
r(1.5%rdg + 5dgt)
250VDC/ACrms
99.99Hz
0.01Hz
999.9Hz 0.1Hz
r(1.2%rdg + 3dgt)
9.999kHz
0.001kHz
99.99kHz 0.01kHz
999.9kHz
0.1kHz
9.999MHz 0.001MHz
r(1.5%rdg + 4dgt)
10.00MHz
0.01MHz
Sensitivity: >8Vrms
In AC voltage frequency range: 10Hz ÷10kHz, sensitivity >15Vrms
Duty Cycle
Range (*) Resolution Accuracy
0.5% y 99%
0.1%
r(1.2%rdg + 2dgt)
Pulse frequency range: 5Hz y 10kHz, Pulse duration: 100Ps y 100ms
In AC voltage frequency range: 10Hz ÷10kHz, sensitivity >15Vrms