13. Step the INCREASE (
) pushbutton to scroll through the setup modes until you return to
RUN.
14. Press ENTER (
). At this point, the BarGraph will return to normal operation.
15. Recheck your BarGraph's calibration values with the lab-standard input signal source and
verify that you have accomplished the desired results. Values between each of the calibra-
tion points should be checked to insure that accuracy is maintained across the full range of
calibration values.