EasyManua.ls Logo

HunterLab LabScan XE - Cleaning the Labscan XE

HunterLab LabScan XE
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LabScan XE Supplemental Manual for EasyMatch QC
18
The green tile reading should vary by no more than
0.15 XYZ units from the values given on the tile. If
your reading is out of this specification, first clean the standard tiles. Then standardize and measure
the green tile again. If the reading still does not meet this specification, contact HunterLab Technical
Support for assistance. Please read “When You Need Assistance” prior to contacting HunterLab.
Cleaning the LabScan XE
The LabScan XE is NOT waterproof, but you may wipe the exterior of the case with a damp cloth.
Also remove the port plate and clean the glass underneath with the included lens brush or with a small
amount of soapy water on a lint-free cloth or towel.

Other manuals for HunterLab LabScan XE

Related product manuals