01/03 Installation 3-1
Installation
The following instructions guide you through the initial installation of your LabScan XE system.
1. Unpack all cartons and remove wrappings and cable ties. Inspect for damage and notify the carrier
and HunterLab immediately if any is discovered. Save the packing materials in case it becomes
necessary to return the instrument to the factory.
2. Place the instrument on a flat working surface where the measurements will be made. Place the
computer in close proximity to the sensor.
3. Ensure that the on/off switch on the back of the sensor is set to “off.”
4. Connect the power cord to the sensor and plug it into a power outlet.
Note: Refer to the Specifications chapter for recommendations concerning the power line and its
conditioning.
CAUTION
Use only the power cord included with this instrument or a replacement obtained
from HunterLab (HunterLab Part Number 10-0000-55 for 110V, A13-1002-656 for
220V). Be certain that the power cord is in good condition before connecting it.
The LabScan XE is grounded using the grounding portion of this power cord. Only
plug this cord into a properly grounded power outlet. Do not use an inappropriate
adapter to plug the instrument into an ungrounded outlet or electric shock may
occur. More information on the wiring of the power cord can be found in the
Specifications chapter.
5. Connect the female end of the 9-pin communications cable to the computer. Connect the male end
of the 9-pin cable to the sensor.
6. Remove the tape covering the sample port of the instrument.
7. Place the standard (2-inch) port plate at the sample port and snap it into place. Cover the sample
port with a spare tile or the sample clamp to prevent dust from accumulating inside the instrument.
Do so whenever the instrument is not in use.
8. Turn on the sensor by switching the on/off button on the back of the sensor to the “on” position.