    Ashley MahoneySep 23, 2025
    What to do if Huntleigh Monitor shows 'NIBP self-test error'?
    • J
      James CoxSep 23, 2025
      If the Huntleigh Monitor displays 'NIBP self-test error', it means there is a problem with the sensor or other hardware. Do not use the NIBP module and contact service personnel.
  • W
    William WheelerAug 25, 2025
    What to do if Huntleigh SC300 Monitor displays 'Weak signal'?
    • E
      Edgar HamiltonAug 25, 2025
      If your Huntleigh Monitor shows 'Weak signal', check the patient’s vital signs and consider changing the measurement site.
  • C
    carla52Aug 25, 2025
    How to fix 'Loose cuff' error on Huntleigh Monitor?
    • R
      Rose ThomasAug 26, 2025
      If the Huntleigh Monitor displays 'Loose cuff', it indicates that the NIBP cuff is not properly connected. To resolve this, check the cuff and connect it again properly.
  • R
    Ronald CrossAug 27, 2025
    What to do if Huntleigh Monitor shows 'Air leak'?
    • D
      diana46Aug 27, 2025
      If the Huntleigh Monitor indicates 'Air leak', it means the NIBP cuff isn't properly connected, or there's a leak in the airway. Check the connection, and if needed, use a new cuff. If the problem continues, contact service personnel.
  • T
    Tina WilliamsSep 23, 2025
    What does 'NIBP communication error' mean on Huntleigh SC300?
    • V
      Virginia HoganSep 23, 2025
      If the Huntleigh Monitor displays 'NIBP communication error', it means there is a problem with the NIBP module or host. Restart the monitor, if the error continues, contact service personnel.
  • J
    Jessica LivingstonSep 13, 2025
    What should I do if Huntleigh SC300 Monitor shows 'NIBP over voltage protective'?
    • A
      Alexandra MeltonSep 13, 2025
      If the Huntleigh Monitor displays 'NIBP over voltage protective', the airway may be occluded. Check the airway and the patient’s condition. If the error continues, contact service personnel.
  • W
    williamsmarkSep 12, 2025
    How to fix 'Ear thermometer communication error' on Huntleigh SC300?
    • J
      james63Sep 12, 2025
      If the Huntleigh Monitor displays 'Ear thermometer communication error', it means the battery is too low or there is a problem with the communication module. Replace the battery and restart the monitor. If the error continues, contact service personnel.
  • H
    hayneswalterSep 12, 2025
    What does 'NIBP air leak' mean on Huntleigh Monitor?
    • S
      Sherri HillSep 12, 2025
      If the Huntleigh Monitor displays 'NIBP air leak', it means there is a leak in the airway. Check and replace the parts that cause the leak. If the error continues, contact service personnel.
  • W
    wmilesSep 4, 2025
    How to fix 'NIBP weak signal' error on Huntleigh SC300 Monitor?
    • J
      Jacob JohnsonSep 4, 2025
      If the Huntleigh Monitor displays 'NIBP weak signal', it means that the cuff is loose or the signal is weak. Check the patient type setting and the connection, or replace the cuff. If the error continues, contact service personnel.
  • V
    Victoria DouglasSep 2, 2025
    How to troubleshoot 'Air pressure Err' on Huntleigh SC300 Monitor?
    • D
      dsteeleSep 2, 2025
      If the Huntleigh Monitor displays 'Air pressure Err', it means the pressure is not stable, possibly due to hose entanglement. Check the connection or use a new cuff. If the problem continues, contact service personnel.

Summary

General Safety Information

Warnings and Cautions

Provides critical safety warnings and cautions for operating the monitor.

Description of Controls, Indicators, Symbols and Displays

Front Panel Controls and Symbols

Identifies and explains the function of front panel controls and symbols.

Rear Panel Controls Identification

Details the controls and connectors located on the rear panel of the monitor.

Side Panel Controls Identification

Identifies controls and optional components on the side panels.

Symbols and Indicators Explained

Explains the meaning of various symbols and indicators used on the monitor.

Description of Monitor Controls

Details the functions of the monitor's operational keys and buttons.

Setting up the Monitor

Power Cable Connections

Explains how to connect the AC power cord to the monitor and a power outlet.

Connecting Monitor Accessories

Details how to connect NIBP hoses, cuffs, and SpO2 sensors to the monitor.

Battery Operation

Using the Monitor

Power On and Self-Test (POST)

Explains the power-on self-test process and what to verify.

NIBP Monitoring Procedures

Starting and Stopping NIBP Measurements

Describes how to initiate and terminate NIBP measurements.

Automatic NIBP Measurements

Explains how to configure and use automatic NIBP measurement intervals.

SpO2 and Pulse Rate Monitoring

Temperature Monitoring

Pairing Thermometer and Main Unit

Guides on how to establish a wireless connection between the thermometer and monitor.

Temperature Measurement Procedure

Details the procedure for taking a temperature measurement.

Maintenance and Care

Troubleshooting Guide

General Troubleshooting Information

Introduces troubleshooting by explaining error codes and fault presentation.

Corrective Action for Errors

Lists common errors and provides suggestions for correction.

Obtaining Technical Assistance

Guides users on how to contact technical support and obtain service manuals.

Electromagnetic Compatibility

Product Specifications

Huntleigh SC300 Specifications

General IconGeneral
TypeFetal Monitor
ModelSC300
ManufacturerHuntleigh
DisplayLCD
Paper Speed1, 2, 3 cm/min
Parameters MonitoredFetal Heart Rate, Uterine Contractions
Frequency Range/Ultrasound Frequency1 MHz
FHR1 Range30 – 240 bpm
FHR2 Range30 – 240 bpm
UC Range0 to 100 mmHg
TOCO Range0 to 100 mmHg