10
AUTO
-
LENSMETER
2.2. How to Set Origin
The steps of setting origin are as follows:
A. Press the ‘LENS’ (Key 1) button.
B. Press the ‘MEASURE (Key 2) button.
C. Determine the lens to measure by using two -INDEX’, ’+INDEX buttons.
D. ‘D-0.01’ button and ‘D+0.01’ button by the given value in the Standard Lens-set
specification.
E. To save and exit, press the ‘SAVE’ button.
F. Press the ‘EXIT button again. And check if the diopter that has been set up in the
basic display screen is shown.
‘SAVE SCREEN INFORMATION’ is as follows:
CEN: The x, y, coordinates of big center point.
LEN: Average distance between of four points.
CON: The x, y, coordinates of four big points.
DIP: Saving order and the saved diopter.
Buttons for Selecting
Diopter
Buttons for Calibrate
Diopter
SAVE button
Previous screen
button