Handpiece Cleaning
The HydraFacial
handpiece tubing chambers must be cleaned and disinfected using the cleaning brush.
1. On the screen, select Menu. Then, select Handpiece Rinse from the System Rinses.
2. Open the control valve on the handpiece.
3. Detach the handpiece from all tubing by gently twisting the knobs counter-clockwise.
4. Insert the cleaning brush into the vacuum (smaller) opening on the face of the handpiece. Push the brush in and out fully while
rotating three times. Repeat on the blue twist lock connector opening on the back of the handpiece.
5. Wash the brush bristles in soapy water between each handpiece opening insertion to clean off debris.
Optional: Insert the compressed air dispenser into the blue twist lock connector and administer two sprays.
6. Clean the entire outer surface of the handpiece with a germicidal wipe, removing all visible soil. Use a second germicidal wipe
to thoroughly wet the surface of the handpiece and ensure the handpiece remains visibly wet for 2 minutes.
Note: Barbicide is a recommended disinfectant (follow Barbicide’s instructions for use). Isopropyl alcohol is not suitable
as a disinfectant.
7. Re-attach the tubing and cleaning cap to the handpiece.
8. Select Continue To Rinse on the screen to run a system rinse and follow the on-screen prompts.