Technical Specications
Parameter Specication 220–240 VAC Specication 110 120 VAC
7000079 7000078
Voltage 220–240 VAC, 50/60 Hz, nominal 2A, 8A max 100120 VAC, 50/60 Hz, nominal 2A, 8A max
Power Consumption 142 watts, maximum 142 watts, maximum
Mode of Operation Continuous Continuous
Audio Volume Under 70 dB Sound Pressure Level at 1 meter Under 70 dB Sound Pressure Level at 1 meter
Visual Indicators LCD display with system messages LCD display with system messages
Instrument Maker/Model HydraFacial, LLC / HydraFacial Syndeo
HydraFacial, LLC / HydraFacial Syndeo
Dimensions 117 cm / 46 in (H) x 50.8 cm / 20 in (W) x 45.8 cm /18 in (D) 117 cm / 46 in (H) x 50.8 cm / 20 in (W) x 45.8 cm /18 in (D)
Weight 30 kg (65 lbs) 30 kg (65 lbs)
Operating Altitude Maximum of 2000 m Maximum of 2000 m
System Controls Touchscreen monitor, buttons Touchscreen monitor, buttons
Interconnects Custom cables Custom cables
Ingress of Fluids Not protected against the ingress of uid. IPX0 Not protected against the ingress of uid. IPX0
Flammability Equipment not suitable for use in the presence of ammable
anesthetic mixture with air or with oxygen or with nitrous oxide.
Equipment not suitable for use in the presence of ammable
anesthetic mixture with air or with oxygen or with nitrous oxide.
Electrical Conformity This medical equipment has passed all required testing for
electric shock, re and mechanical hazards in accordance
with IEC 60601–1, IEC/EN 60601–1.
This medical equipment has passed all required testing for
electric shock, re and mechanical hazards in accordance with
IEC 60601–1, IEC/EN 60601–1.
Electrical Shock Protection – Classication Class I Class I
Electrical Shock Protection – Degree Type BF Equipment Type BF Equipment
Product Life The Console is a non-sterile, durable good. The System
warranty runs for a period of two (2) years from the date of
installation of the System to the original Customer.
The Console is a non-sterile, durable good. The System
warranty runs for a period of two (2) years from the date of
installation of the System to the original Customer.
Applied Part The Handpiece is the client applied part The Handpiece is the client applied part
Separation From Power Mains Power switch on rear panel provides full separation from
power mains.
Power switch on rear panel provides full separation from
power mains.
Fuse Slow-blow fuse. 250V 8A Slow-blow fuse. 250V 8A
Environmental Conditions
Parameter Specication
For Shipping & Storage
Temperature: Normal (–20 to 60 C)
Relative Humidity: 10—90% RH
Atmospheric Pressure: 500—1060 hPa
Operating Conditions
Temperature: Normal (10 to 40 C)
Relative Humidity: 30—75% RH