10
NOTE: SureTrend software is required for sections 3.1 to 3.3. For
further details, refer to the separate SureTrend User's Manual
located on the SureTrend software CD that is packaged with the
luminometer.
3.1 User ID
The luminometer can store a list of 201 User IDs (User: 0 to 200).
Each test result will be tagged with the currently selected User ID.
Select the User ID menu option, then use the and buttons to
scroll through the list of User IDs, then press the button to select
a User ID, or the button to cancel the selection.
TIP: User:0 is the default selection if no User ID list has been
defined.
3.2 Setting up Program Test Points and Pass/Fail Limits
The luminometer has the ability to store a list of 5001 unique
Program test points (Prog# 0 to 5000), each with its own upper ()
and lower (
) pass/fail result threshold values.
Program test point location names can only be programmed into the
luminometer through the SureTrend software. Program pass/fail
thresholds can either be set-up on the luminometer, or set using the
SureTrend software. If a location name is entered and threshold set
in SureTrend, and then downloaded to the luminometer, the
threshold values can’t be changed via the luminometer. This is a
security feature and prevents inadvertent changes to the thresholds
after being set in SureTrend.
TIP: Prog:0 has a special function, whereby the threshold values
defined for Prog:0 are used as the initial values when setting-up
the other Prog thresholds. Setting Prog:0 with your typical
threshold values will save time when setting-up the other Programs.
The luminometer comes with pass/fail thresholds of 10 RLU for pass
and 30 RLU for fail. These pass/fail thresholds were determined
based on a comprehensive study of ATP, food residue, and
microorganism samples. To receive the report or more information
on setting up pass/fail thresholds contact Hygiena or visit our
Question and Answer section at www.hygiena.com.
To set the Program result thresholds on the luminometer, first select
the PROGRAM menu option by pressing . Scroll to “PROGRAM”
and press
. Then use the and buttons to scroll through the
list of Programs, followed by the button to select the Program to
be changed: