19
Instruction Manual C100-AE
tol.LEd In this menu flashing of tolerance points when measured values are out of tolerance is enabled
or disabled.
Flash = The tolerance points are flashing if measured values are out of the green area.
Lit = Tolerance points do not flash.
PASS.Cd. For protection of programme settings in this menu you can set a password option on or off.
If the password option has been set to "on" the numeric display will force the operator to input a
password consisting of a 6-digit number.
Notice : If the Programming menu is called later by the PRG/CAL key the column
gauge will prompt you to introduce the 6-digit password.
( If you have forgotten the password, you can nevertheless access the
programming menu with the password 240355 ).
FiLtEr In this menu you can select a digital filter for the measured values.
oFF - Digital filter is off.
1...6 - Filter levels
Notice : For normal measurements the filter time constant should be set to 1 or 2.
Higher filter values are necessary only for special applications.