Appendix 3
Diagrams of the sterilization programs STE-23-D Pro
The time required for sterilizer to be ready for routine use after the power is switched is less than
5 minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C
Programs
Temperature (°C)
Pressure (kPa)
Holding time
(min)
Total time (min)
Type
Max load (kg)
Max load per tray
(kg)
Solid 134
134
210
4
15-30
Unwrapped solid material
5.00
1.50
Wrapped
134
134
210
4
38-57
Unwrapped solid material
5.00
1.50
Single-wrapped solid or hollow
material
4.50
1.20
Wrapped
121
121
110
20
46~62
Unwrapped solid material
5.00
1.50
Single-wrapped solid or hollow
material
4.50
1.20
Prion134
134
210
18
45-60
Unwrapped porous material
1.25
0.40
Single-wrapped porous
material
1.10
0.30
Dual-wrapped porous material
0.75
0.25
Single-wrapped hollow
material
4.00
1.25
Dual-wrapped solid and hollow
material
2.00
0.60
Drying
(optional)
1-20
B&D test
134
210
3.5
28-35
Helix test
134
210
3.5
28-35
Vacuum test
20-25