Note: If the cycle is manually interrupted before it reaches the drying phase, the items inside the sterilizer
must be considered not sterile. N20 will appear on the screen. (see Error code description).
Caution: Depending on the phase of the cycle, steam and water can escape from the
sterilization chamber when you open the door.
Tes t Pro gra m s
Helix test
Put the Helix test device into the chamber, then close the door.
Select “Program” from the main menu by pressing then to enter in the menu, select
“Helix test”; in the screen will appear the information as temperature and sterilization time
(holding time), also the date, time, current pressure and current temperature.
Press to start the cycle. The stage, conditions and the status of the cycle will appear on the
display. The sterilizers will perform the program automatically. (see appendix 2).
After finishing the cycle, you may check the indicator and evaluate the result according with the
instructions of the test manufacturer.
B&D test
Put the B&D test package into the chamber, then close the door.
Select “Program” from the main menu by pressing then to enter in the menu, select B&D
test, in the screen will appear the information as the temperature and sterilization time
(holding time), also the date, time, current pressure and current temperature.
Press to start the cycle. The stage, conditions and the status of the cycle will appear on the
display. The sterilizers will perform the program automatically. (see appendix 2).
After finishing the cycle, you may check the indicator and evaluate the result according with the
instructions of the test manufacturer.
Vacuum Test
Select “Program” from the main menu by pressing then to enter in the menu, select
“Vacuum test”.
After closing the door, press to start the cycle. The stage, conditions and the status of the cycle
will appear on the display. The sterilizers will perform the program automatically. (see appendix 2).
In compliance with EN 13060, the test requires that the air leakage rate less than or equal to 0.13
kPa/min. during 10 minutes.
If leakage rate is not greater 0.13, it will show Success.
If the temperature difference between the max. Temperature and the Min. is above 3°C, it will show
void. That means the result of the test is fail. You need run the vacuum test again after the chamber has
cooled down.