3 Installation
General conditions
§ Position the device on a plane surface with a minimum capacity of 60 kgs.
§ The sterilizer should be placed on a level worktable.
§ Leave at least 10 cm between the device rear part and the wall. The clearance required to open
the door is 40cm.
§ Position the sterilizer at such a height as to make it possible for the operator to check the whole
sterilization chamber and carry out the normal cleaning operations.
§ The room where the device is installed must be sufficiently ventilated.
§ Do not install the device near washing basins, taps, etc. where it is likely to be splashed.
§ Do not lean on the door when it is opened.
§ Do not place trays, papers, fluid containers or other objects on the sterilizer.
Power supply connection
Check the label on the back panel of sterilizer to verify the voltage rating for the unit. Failure to connect
the sterilizer to an appropriate power supply could result in damage to the unit and electrical shock to
personnel.
Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only used
for the sterilizer is recommended. Never connect the device pin to reductions of any type.
Location requirements and dimensions (mm)