21
Appendix 2
Diagrams of the sterilization programs
The time required for sterilizer to be ready for routine use after the power is switched is less than 15
minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C
Programs
(STE-8-D)
Temperature (°C)
Pressure (kPa)
Holding time (min)
Total time
(min)
Type
Max load (kg)
Max load per tray
(kg)
SOLID
134
210
4
15-20
Unwrapped solid material
2.00
0.60
121
110
20
30-35
WRAPPED
134
210
4
25-40
Unwrapped solid material
2.00
0.60
121
110
20
45-50
Single-wrapped solid or
hollow material
1.50
0.50
TEXTILE
134
210
8
30-45
Unwrapped porous material
0.50
0.15
Single-wrapped porous
material
0.35
0.10
121
110
30
55-60
Dual-wrapped porous material
0.25
0.10
Single-wrapped hollow
material
1.50
0.30
Dual-wrapped solid and
hollow material
1.00
0.30
PRION
134
210
18
40-55
Unwrapped porous material
0.50
0.15
Single-wrapped porous
material
0.35
0.10
Dual-wrapped porous material
0.25
0.10
Single-wrapped hollow
material
1.50
0.30
Dual-wrapped solid and
hollow material
1.00
0.30
LIQUID
(Optional)
134
210
10
35-55
Liquid
0.60
0.20
121
110
30
40-50
Drying
(Optional)
1-20
B&D test
134
210
3.5
10-20
Helix test
134
210
3.5
10-20
Vacuum test
17-20