25
The time required for sterilizer to be ready for routine use after the power is switched is less than 15
minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C
Programs
(STE-29-D)
Temperature (°C)
Pressure (kPa)
Holding time (min)
Total time (min)
Type
Max load (kg)
Max load per tray
(kg)
SOLID
134
210
4
30-45
Unwrapped solid material
6.0
2.0
WRAPPED
134
210
8
45-60
Unwrapped porous material
3.5
2.0
Single-wrapped porous
material
3.0
2.0
Dual-wrapped porous material
2.5
4.0
2.0
2.0
121
110
30
50-65
Single-wrapped hollow
material
3.5
2.0
Dual-wrapped solid and
hollow
material
6.0
2.0
PRION
134
210
18
55-70
Unwrapped porous material
3.5
2.0
Single-wrapped porous
material
3.0
2.0
Dual-wrapped porous material
2.5
2.0
Single-wrapped hollow
material
4.0
2.0
Dual-wrapped solid and
hollow
material
3.5
2.0
LIQUID
(optional)
121
110
30
55-70
Liquid
3.5
2.0
Drying
(optional)
1-20
B&D test
134
210
3.5
35-40
Helix test
134
210
3.5
35-40
Vacuum test
20-25