3. Remove the base from the analyzer. 4. Clean the entire base using a lint-free
wipe or cotton swab dampened with 70%
isopropyl alcohol, mild liquid soap and water,
or a 10% bleach solution.
reflectors
lens
activator
axle
5. Clean the 3 white reflectors, the axle, and the
activator using a new lint-free wipe or cotton
swab dampened with 70% isopropyl alcohol,
mild liquid soap and water, or a 10% bleach
solution. DO NOT clean the lens.
6. After the base has dried, reattach it to the
analyzer and ensure it is locked into place.
Maintenance and Troubleshooting