3 Functions and features
The unit measures and monitors the system pressure in a plant�
3.1 Applications
Type of pressure: relative pressure
Order no. Measuring range
Permissible
overpressure
Bursting
pressure
bar PSI bar PSI bar PSI
PI2793 -1���25 -14�4���362�7 100 1450 350 5075
PI2794 -1���10 -14�5���145 50 725 150 2175
PI2795 -1���4 -14�5���58 30 435 100 1450
PI2796 -0�124���2�5 -1�8���36�27 20 290 50 725
PI2797 -0�05���1 -0�73���14�5 10 145 30 435
PI2799 -1���1 -14�5���14�5 10 145 30 435
mbar inH2O bar inH2O bar inH2O
PI2798 -12�4���250 -5�0��100�4 10 4015 30 12044
PI2789 -5���100 -2�00���40�16 4 1606 30 12044
Avoid static and dynamic overpressure exceeding the given overload
pressure by taking appropriate measures�
The indicated bursting pressure must not be exceeded�
Even if the bursting pressure is exceeded only for a short time, the unit
may be destroyed� ATTENTION: Risk of injury!
Not suitable for use where the criteria for paragraph D10�1�2/74-03 of the 3A
standard 74-03 have to be met�
4 Function
• The unit displays the current system pressure�
• It generates output signals according to the operating mode and the parameter
setting�
• It moreover provides the process data via IO-Link�
• The unit is laid out for fully bidirectional communication� So, the following
options are possible:
- Remote display: reading and display of the current system pressure�
- Remote parameter setting: reading and changing the current parameter
setting�