15
Maintenance and cleaning
The device is maintenance-free. It is only subject to the
natural wear and tear of components and their statistical
failure rate.
Cleaning:
For cleaning disconnect the mains
plug!
Use only cleaning agents which have been approved by
IKA
®
to clean IKA
®
devices.
Dirt Cleaning agent
Dye Isopropyl alcohol
Construction material
Water containing tenside/ isopropyl alcohol
Cosmetics
Water containing tenside/ isopropyl alcohol
Foodstuff Water containing tenside
Fuel Water containing tenside
For materials which are not listed, please request
information from IKA
®
application support.
Wear protective gloves while cleaning the device.
Electrical devices may not be placed in the cleansing agent
for the purpose of cleaning.
Do not allow moisture to get into the device when cleaning.
Before using another than the recommended method for
cleaning or decontamination, the user must ascertain with
IKA
®
that this method does not damage the device.
Spare parts order:
When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
-
item and designation of the spare part,
see www.ika.com, spare parts diagram and list
- Software version.
Repair:
Please send the device for repair only after it has
been cleaned and is free from any materials which
may constitute a health hazard.
For repair, please request the “Decontamination Cer-
tificate” from IKA
®
, or download printout of it from the
IKA
®
website www.ika.com.
If you require servicing, return the device in its original
packaging. Storage packaging is not sucient. Please also
use suitable transport packaging.
Error codes
Any malfunctions during operation will be identified by an error message on the display.
Proceed as follows in such cases:
)
Switch off device using the main switch at the back of the device
)
Carry out corrective measures
)
Restart device
Error code Effect Cause Solution
Err 02 Pump off
Heating off
Motor over current (rate current) - Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Err 03 Pump off
Heating off
Motor over current (Max. current) - Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Err 04 Pump off
Heating off
Motor hall signal missing - Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Err 05 Pump off
Heating off
Too high liquid level - Check the liquid level and buoyage
Err 06 Pump off
Heating off
Too low liquid level - Check the liquid level and buoyage
Err 07 Pump off
Heating off
Too high voltage - Check the mains power
Err 08 Pump off
Heating off
Too low voltage - Check the mains power